302 Chapter 6
Replacement Procedure
Receiver Module Replacement for 4-Port model
Receiver Module Replacement for 4-Port model
Tools Required
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end wrench, 5/8 inch
Open-end torque wrench, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA
connector)
Torque driver set, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting SMA
connector)
Removal Procedure
Refer to Figure 6-18 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266.
Step 2. Remove the front panel as described in “Front Panel Removal” on page 267.
Step 3. Disconnect the two semi-rigid cable (item 1).
Step 4. Disconnect the two semi-rigid cable (item 2).
Step 5. Disconnect the four semi-rigid cables (item 3).
Step 6. Remove the four TORX T10 screws fastening the module to the bottom. (item 4)
Step 7. Slide the receiver module (item 5).