Chapter 2 65
Performance Test
E5071C Performance Test Record
2. Performance Test
RF Output Level Linearity Test (@ Port1)
1. Power Sensor: E9304A Option H18
CW Frequency: 9 kHz (relative to 0 dBm reference)
Option 230, 240, 280, 430, 440, 480
Power level
[dBm]
Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-20 ± 0.75 ± 0.23
-17.5 ± 0.75 ± 0.23
-15 ± 0.75 ± 0.23
-12.5 ± 0.75 ± 0.23
-10 ± 0.75 ± 0.23
-7.5 ± 0.75 ± 0.22
-5 ± 0.75 ± 0.22
-2.5 ± 0.75 ± 0.22
2.5 ± 0.75 ± 0.21
5 ± 0.75 ± 0.21
7.5 ± 0.75 ± 0.21
10 ± 0.75 ± 0.21
CW Frequency: 100 kHz (relative to 0 dBm reference)
8.025 G ± 1.0 ± 0.30
8.5 G ± 1.0 ± 0.30
Frequency [Hz] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Power level
[dBm]
Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-20 ± 0.75 ± 0.23
-17.5 ± 0.75 ± 0.23
-15 ± 0.75 ± 0.23
-12.5 ± 0.75 ± 0.23
-10 ± 0.75 ± 0.23