EasyManuals Logo

Agilent Technologies E5071C Service Guide

Agilent Technologies E5071C
385 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #284 background imageLoading...
Page #284 background image
284 Chapter 6
Replacement Procedure
Digital Bridge Board Replacement
Digital Bridge Board Replacement
Tools Required
TORX screwdriver, T10, T15, and T20
Removal Procedure
Refer to Figure 6-12 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266.
Step 2. Remove the three TORX T10 screws (item 1) fastening the angle.
Step 3. Disconnect the flat cables (item 2) from the digital bridge board.
Step 4. Remove the two TORX T10 screws (item 3) fastening the board to the rear panel.
Step 5. Lift the digital bridge board (item 4).
Step 6. Disconnect the cables (item 5) from the digital bridge board.
Step 7. Remove the two TORX T10 screw (item 6) fastening the angle.

Table of Contents