Chapter 4 199
Troubleshooting
Function Specific Troubleshooting
4. Troubleshooting
• Check the cable between the touch screen controller and the serial interface connector
on the CPU Mother Board Assembly.
• If no problem is found in the above checks, a failure in the CPU Mother Board
Assembly is suspected.
To Check the LCD
Procedure
Step 1. Press key.
Step 2. Click Service Menu and, then, Test Menu in the softkeys menu.
Step 3. Click Display in the test menu. The whole of the LCD screen turns Red, Green, Blue,
White and Black every 2 seconds and returns to the measurement view. If the color test
screen doesn't appear correctly, perform step 4.
Step 4. Connect an external VGA monitor to the VIDEO output port on the E5071C rear panel.
• If the monitor screen view is the same as the LCD display, the problem seems in the
PCA PCI DSP Card. Check first if the PCA PCI DSP Card is securely connected to the
CPU Mother Board Assembly.
• If only the LCD display has a problem, check the flat cable between the PCA Front
Key and PCA Digital Bridge.
• If the LCD is not illuminated with backlight, check the Inverter Unit Assembly and the
cable between the Inverter Unit Assembly and PCA Front Key. Also check the cables
between the LCD-TPANEL Assembly and Inverter Unit Assembly.
• If the cables are normal, check the LCD-TPANEL Assembly.
To Check the External Keyboard
Procedure
Step 1. Connect the external keyboard to the E5071C rear panel USB connector.
Step 2. Turn the instrument on.
Step 3. Press key.
Step 4. Press and keys on the external keyboard, and verify that the cursor on the menu
bar moves up and down. If it doesn't work, the external keyboard or the CPU Mother Board
Assembly may be faulty.
To Check the Mouse
Procedure
Step 1. Connect the mouse to the E5071C rear panel USB connector.
Step 2. Turn the instrument on.
Step 3. Move the mouse and verify that the mouse pointer moves smoothly. If it doesn't move

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals