294 Chapter 6
Replacement Procedure
Analog Base Module Replacement
Analog Base Module Replacement
Tools Required
TORX screwdriver, T10, T15, and T20
Open-end torque wrench, 5/8 inch (set to 1.97 N-m / 17.4 lb-in)
Removal Procedure
Refer to Figure 6-16 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266
Step 2. Remove the Power Supply as described in “Power Supply Assembly Replacement” on
page 282
Step 3. Remove the two TORX T10 screws (item 1) fastening the plate and lift it.
Step 4. Remove the four 5/8 inch nut (item 2) fastening the rear panel.
Step 5. Disconnect the flat cable (item 4) from analog motherboard.
Step 6. Remove the three TORX T10 screws (item 3) fastening the analog base module.
Step 7. Push the analog base module (item 5) till the end of the wall. Carefully lift it when the
height of BNC connectors are higher than rear chassis wall.
Step 8. Remove the spacer (item 6) from analog base module.

Table of Contents

Other manuals for Agilent Technologies E5071C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E5071C and is the answer not in the manual?

Agilent Technologies E5071C Specifications

General IconGeneral
BrandAgilent Technologies
ModelE5071C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals