Chapter 3 173
Adjustment
Required Adjustment after Replacing Assembly
1. Chapter Title
2. Chapter Title 3. Adjustment
4. Chapter Title
5. Chapter Title
Virtual Bridge Coefficient Adjustment
This item writes the coefficient factors to the E5071C.
Required equipment for the source power adjustment
None
Source Output Power Adjustment
The purpose of this procedure is to adjust the variable attenuator in Level Vernier board,
and the gains among Level Vernier, Distributor and Receiver.
Required equipment for the test port characteristics adjustment
Description Recommended Model
Power Meter Agilent E4419A/B
Power Sensor Agilent E9304A Opt H18
Receiver IF Range Adjustment
The purpose of this procedure is to adjust IF Range.
Required equipment for the test port characteristics adjustment
Description Recommended Model
Calibration Kit Agilent 85032F
N(m)-N(m) Cable Agilent N6314A (p/n 8120-8862)
Receiver Ports Characteristics Adjustment
The purpose of this procedure is to adjust source match, directivity and tracking.
Required equipment for the test port characteristics adjustment
Description Recommended Model
Calibration Kit Agilent 85032F
N(m)-N(m) Cable Agilent N6314A (p/n 8120-8862)
Receiver Absolute Gain Adjustment
This item writes the coefficient in Absolute measurement.
Required equipment for the test port characteristics adjustment
None