282 Chapter 6
Replacement Procedure
Power Supply Assembly Replacement
Power Supply Assembly Replacement
Tools Required
TORX screwdriver, T10, T15, and T20
Torque screwdriver, TORX T10 (set to 0.79 N-m / 7 lb-in)
Removal Procedure
Refer to Figure 6-11 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 266.
Step 2. Remove the two TORX T10 screws (item 1) fasting the angle (item 2) and lift it.
Step 3. Disconnect the power supply cables (item 3).
Step 4. Disconnect the cable form analog motherboard(item 4).
Step 5. Remove the two TORX T10 screws (item 5) from the rear panel.
Step 6. Remove the two TORX T10 screws and washers (item 6) fastening the power supply
holder.
Step 7. Lift the power supply assembly toward front side.
Step 8. Remove the three TORX T10 screws (item 7) fasting the angle.