60 Chapter 2
Performance Test
E5071C Performance Test Record
RF Output Level Accuracy and Flatness Test
2. Power Sensor: E9304A Option H19
Level Accuracy Test (@ Port 1)
Power level
[dBm]
Frequency
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
0 50 M ± 0.65 ± 0.19
Level Flatness Test (@ Port 1) (at 0 dBm, relative to 50 MHz reference)
Option 230, 430
3.025 G ± 1.0 ± 0.24
3.525 G ± 1.0 ± 0.24
4.025 G ± 1.0 ± 0.24
4.525 G ± 1.0 ± 0.24
5.025 G ± 1.0 ± 0.24
5.525 G ± 1.0 ± 0.24
6.025 G ± 1.0 ± 0.28
6.525 G ± 1.0 ± 0.28
7.025 G ± 1.0 ± 0.28
7.525 G ± 1.0 ± 0.28
8.025 G ± 1.0 ± 0.28
8.5 G ± 1.0 ± 0.28
Frequency [Hz] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Frequency [Hz] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
9 k ± 1.0 ± 0.28
20 k ± 1.0 ± 0.28
70 k ± 1.0 ± 0.28
100 k ± 1.0 ± 0.28