E5071C Performance Test Record
2. Performance Test
Dynamic Accuracy Test
@ Reference power level: -10 dBm
Direction: S21
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.010
5 ± 0.075 ± 0.010
0 ± 0.041 ± 0.007
-5 ± 0.031 ± 0.009
-15 ± 0.030 ± 0.006
-20 ± 0.035 ± 0.007
-25 ± 0.040 ± 0.010
-30 ± 0.045 ± 0.011
-35 ± 0.056 ± 0.015
-40 ± 0.067 ± 0.019
-45 ± 0.078 ± 0.021
-50 ± 0.091 ± 0.021
-55 ± 0.106 ± 0.024
-60 ± 0.125 ± 0.024
-65 ± 0.151 ± 0.026
-70 ± 0.189 ± 0.025
-75 ± 0.248 ± 0.011
-80 ± 0.346 ± 0.014
-85 ± 0.509 ± 0.021
-90 ± 0.785 ± 0.033
-95 ± 1.248 ± 0.069
-100 ±2.008 ± 0.060
Direction: S12
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
10 ± 0.207 ± 0.010
5 ± 0.075 ± 0.010