11 0 Chapter 2
Performance Test
E5071C Performance Test Record
Option 235
Direction: S21
Frequency
range [Hz]
IF bandwidth
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
100 k - 300 k 10 < -100 1.73
300 k - 10 M 10 < -110 3.63
10 M - 3 G 10 < -120 4.95
Direction: S12
Frequency
range [Hz]
IF bandwidth
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
100 k - 300 k 10 < -100 1.73
300 k - 10 M 10 < -110 3.63
10 M - 3 G 10 < -120 4.95
Option 240
Direction: S21
Frequency
range [Hz]
IF bandwidth
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
9 k - 100 k 10 < -100 1.73
100 k - 300 k 10 < -100 1.73
300 k - 10 M 10 < -110 3.63
10 M - 3 G 10 < -120 4.95
3 G - 4.5 G 10 < -110 3.32
Direction: S12
Frequency
range [Hz]
IF bandwidth
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
9 k - 100 k 10 < -100 1.73
100 k - 300 k 10 < -100 1.73
300 k - 10 M 10 < -110 3.63
10 M - 3 G 10 < -120 4.95
3 G - 4.5 G 10 < -110 3.32