Chapter 2 121
Performance Test
E5071C Performance Test Record
2. Performance Test
Direction S12
Frequency
range [Hz]
IF bandwidth
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
100 k - 300 k 10 < -97 1.52
300 k - 10 M 10 < -107 1.26
10 M - 3 G 10 < -123 1.24
100 k - 300 k 3 k < -72 1.52
300 k - 10 M 3 k < -82 1.13
10 M - 3 G 3 k < -98 1.08
Option 240
Direction S21
Frequency
range [Hz]
IF bandwidth
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
9 k - 100 k 10 < -97 1.52
100 k - 300 k 10 < -97 1.52
300 k - 10 M 10 < -107 1.26
10 M - 3 G 10 < -123 1.24
3 G - 4.5 G 10 < -123 1.24
9 k - 100 k 3 k < -72 1.52
100 k - 300 k 3 k < -72 1.52
300 k - 10 M 3 k < -82 1.13
10 M - 3 G 3 k < -98 1.08
3 G - 4.5 G 3 k < -98 1.08
100 k - 300 k 3 k < -72 1.52
300 k - 10 M 3 k < -82 1.13
10 M - 3 G 3 k < -98 1.08
Frequency
range [Hz]
IF bandwidth
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]