EasyManua.ls Logo

Agilent Technologies E5071C

Agilent Technologies E5071C
385 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 135
Performance Test
E5071C Performance Test Record
2. Performance Test
Direction S32
Frequency
range [Hz]
IF bandwidth
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
100 k - 300 k 10 < -97 1.52
300 k - 10 M 10 < -107 1.26
10 M - 3 G 10 < -123 1.24
3 G - 5 G 10 < -123 1.24
5 G - 6 G 10 < -123 1.16
6 G - 7 G 10 < -117 1.08
7 G - 8.5 G 10 < -117 1.13
100 k - 300 k 3 k < -72 1.52
300 k - 10 M 3 k < -82 1.13
10 M - 3 G 3 k < -98 1.08
3 G - 5 G 3 k < -98 1.08
5 G - 6 G 3 k < -98 1.08
6 G - 7 G 3 k < -92 1.10
7 G - 8.5 G 3 k < -92 1.10
Direction S43
10 M - 3 G 3 k < -98 1.08
3 G - 5 G 3 k < -98 1.08
5 G - 6 G 3 k < -98 1.08
6 G - 7 G 3 k < -92 1.10
7 G - 8.5 G 3 k < -92 1.10
Frequency
range [Hz]
IF bandwidth
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
Frequency
range [Hz]
IF bandwidth
[Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
100 k - 300 k 10 < -97 1.52
300 k - 10 M 10 < -107 1.26
10 M - 3 G 10 < -123 1.24
3 G - 5 G 10 < -123 1.24

Table of Contents

Other manuals for Agilent Technologies E5071C

Related product manuals