Chapter 2 165
Performance Test
E5071C Performance Test Record
2. Performance Test
Port 3
Load match 6 G - 8.5 G < -10 0.64
Transmission
Tracking
100 k - 300 k ± 1.5 ± 0.03
Transmission
Tracking
300 k - 3 G ± 1.0 ± 0.20
Transmission
Tracking
3 G - 6 G ± 1.0 ± 0.24
Transmission
Tracking
6 G- 8.5 G ± 1.0 ± 0.35
Reflection
Tracking
100 k - 300 k ± 1.5 ± 0.04
Reflection
Tracking
300 k - 3 G ± 1.0 ± 0.04
Reflection
Tracking
3 G- 6 G ± 1.0 ± 0.66
Reflection
Tracking
6 G- 8.5 G ± 1.0 ± 0.64
System
performance
Frequency
range [Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
System
performance
Frequency
range [Hz]
Test limit
[dB]
Test result
[dB]
Measurement
uncertainty [dB]
Directivity 100 k - 300 k < -20 0.09
Directivity 300 k -3 G < -25 0.81
Directivity 3 G - 6 G < -20 0.40
Directivity 6 G - 8.5 G < -15 0.67
Source match 100 k - 300 k < -20 0.05
Source match 300 k -3 G < -25 3.00
Source match 3 G - 6 G < -20 1.94
Source match 6 G - 8.5 G < -15 0.57
Load match 100 k - 300 k < -12 0.04
Load match 300 k - 3 G < -17 0.59
Load match 3 G - 6 G < -12 0.66
Load match 6 G - 8.5 G < -10 0.64