Chapter 4 193
Troubleshooting
Troubleshooting Using Diagnostics Test
4. Troubleshooting
Figure 4-15 Flowchart for troubleshooting on DC-BUS Test
e5071cse2001
DC-BUS TEST
SYN:SRC: FAIL?
SYN:LCL: FAIL?
VNR: FAIL?
ALL 0V?
ALL xxV?
Replace
Analog Base B'd
Sequential Error
Add Receiver B'd
BAS:GND FAIL?
BAS:TEMP FAIL?
BAS:VM9 FAIL?
BAS:OTHER
FAIL?
END
8< room
temperature < 38
Replace
Analog Base B'd
Rec B'd Trouble
Replace
Analog Base B'd
Replace
SRC B'd
Replace
LCL B'd
Replace
VNR B'd
Replace
Analog Base B'd
BAS:VM9 FAIL?
Replace
Analog Base B'd
Room tempareture is
out of condition.
NO
YES
No Rec B'd?
NO
NO
YES
NO
YES
YES
NO
NO
YES
NO
YES
YES
NO
YES
NO
YES
NO
YES
NO
YES
NO
NO
YES