338 Appendix B
System Recovery
System Recovery
Step 6. Select Restore.
Figure B-1 Activate Acronis Startup Recovery Manager
e5071cse1079
Step 7. A confirmation dialog box is displayed. Push the enter to select Yes to start the recovery of
the factory backup. The recovery takes a few minutes depending on the amount of data.
Figure B-2 Recover Factory Backup
e5071cse1034
Step 8. When the recovery is finished, push the enter to select OK. The E5071C restart
automatically.
Step 9. After restart, the screen for initial registration appears. Execute initial registration. For
information on the execution procedure, refer to
“Initial Registration of E5071C” on
page 357.
Step 10. Execute the calibration of the touch screen. For information on the execution procedure,
refer to
“Calibration of the Touch Screen” on page 359.