EasyManua.ls Logo

Agilent Technologies E5071C

Agilent Technologies E5071C
385 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 55
Performance Test
E5071C Performance Test Record
2. Performance Test
With Option 1E5
Frequency [Hz] Test limit [Hz] Test result [Hz] Measurement
uncertainty [Hz]
9 k ± 0.009 ± 0.0003
100 k ± 0.1 ± 0.003
50 M ± 50 ± 0.58
134.1 M ± 134.1 ± 1.6
548 M ± 0.548 k ± 6.4
3 G ± 3 k ± 35
8.5 G ± 8.5 k ± 99
Option 285, 485
Without Option 1E5
Frequency [Hz] Test limit [Hz] Test result [Hz] Measurement
uncertainty [Hz]
100 k ± 0.5 ± 0.004
50 M ± 250 ± 1.7
134.1 M ± 670.5 ± 4.6
548 M ± 2.74 k ± 16.4
3 G ± 15 k ± 90
8.5 G ± 42.5 k ± 254
With Option 1E5
Frequency [Hz] Test limit [Hz] Test result [Hz] Measurement
uncertainty [Hz]
100 k ± 0.1 ± 0.003
50 M ± 50 ± 0.58
134.1 M ± 134.1 ± 1.6
548 M ± 0.548 k ± 6.4
3 G ± 3 k ± 35
8.5 G ± 8.5 k ± 99