Page #71 background imageLoading...
Page #71 background image
Chapter 2 71
Performance Test
E5071C Performance Test Record
2. Performance Test
CW Frequency: 100 kHz (relative to 0 dBm reference)
Power level
[dBm]
Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-20 ± 0.75 ± 0.24
-17.5 ± 0.75 ± 0.24
-15 ± 0.75 ± 0.24
-12.5 ± 0.75 ± 0.24
-10 ± 0.75 ± 0.24
-7.5 ± 0.75 ± 0.23
-5 ± 0.75 ± 0.23
-2.5 ± 0.75 ± 0.23
2.5 ± 0.75 ± 0.22
5 ± 0.75 ± 0.22
7.5 ± 0.75 ± 0.22
10 ± 0.75 ± 0.22
CW Frequency: 50 MHz (relative to 0 dBm reference)
-2.5 ± 0.75 ± 0.23
2.5 ± 0.75 ± 0.22
5 ± 0.75 ± 0.22
7.5 ± 0.75 ± 0.22
10 ± 0.75 ± 0.22
Power level
[dBm]
Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Power level
[dBm]
Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-20 ± 0.75 ± 0.24
-17.5 ± 0.75 ± 0.24
-15 ± 0.75 ± 0.24
-12.5 ± 0.75 ± 0.24
-10 ± 0.75 ± 0.24
-7.5 ± 0.75 ± 0.23