Chapter 4: Instrument QC 71
4 If necessary, click Manual to adjust settings manually.
Detectors/Amps controls are enabled and Target Values appear in a
separate window. Use the controls to set PMTs to ±2 channels of the target
values for fluorescence PMTs, or ±5 channels for SSC.
Click Initial PMT to revert to automatic settings, or
Click Time Delay or Compensation to save settings and proceed.