EasyManua.ls Logo

BD FACSCalibur - Page 73

BD FACSCalibur
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4: Instrument QC 73
If you do not select an option within 5 seconds, the software automatically
proceeds to the Compensation view.
Compensation Adjustment
1 Remove the unlabeled (or unlabeled and APC) bead tube from the
cytometer.
2 Gently mix and install the mixed bead tube.
3 At the Comp view, click Start.
If you proceed with compensation after a failed time-delay calibration, the
compensation values you obtain for FL3–%FL4 and FL4–%FL3 might vary
substantially. If compensation appears incorrect during optimization, verify
time-delay calibration. A failed calibration is noted in the comments section
of the Summary Report.
Running the wrong FL3 beads will result in inaccurate FL4–%FL3 results.
Make sure to install the appropriate mixed beads tube for the assay.

Table of Contents

Related product manuals