92 BD FACSCalibur Instructions for Use
Click the up or down arrow to change the Amp Gain value by .01.
Press the Option key and click the arrow to change the value by 0.1.
To change the value in larger increments, click between the arrows and
drag the slider control.
5 In the Threshold window, select FSC-H as the primary parameter.
6 Adjust the FSC Threshold Value to remove most of the debris without
cutting off the lymphocytes.
9
7 Click Pause, and then Abort in Acquisition Controls.
8 (Optional) Remove the isotype control tube from the SIP.
; Tip Remove the tube to conserve sample.