Calibration
Required Modules for Different Input/Output Signal Combinations
Consider the type of your instrument before doing a calibration:
• What are the input and output quantities/ports?
• Is the input signal measured, generated/simulated, controlled
or keyed (Input method)?
• Is the output signal measured or keyed (Output method)?
Use the following tables find out the required modules for each sup-
ported input/output quantity and method:
Table for choosing the input module Table for choosing the output module
QUANTITY METHOD REQ. MODULE QUANTITY METHOD REQ. MODULE
Voltage
Measured
E Voltage
Measured
E
Generated
ET
Keyed none
Keyed none HART
E
Low Voltage
E measured
E Low Voltage
E Measured
E
ET measured
ET
ET Measured
ET
Generated
ET
Keyed none
Keyed none HART
E
Current
Measured
E Current
Measured
E
E generated
E
Keyed none
ET generated
ET
HART
E
Keyed none
Resistance
Measured
ET
Resistance
Measured
ET
Keyed none
Simulated
ET
HART
E
Keyed none
Pressure
Measured
INT
or
EXT
Pressure
Measured
INT or EXT
Keyed none
Controlled
INT
or
EXT
HART
E
Controlled/Meas none
Temp.
Keyed none
Keyed none HART
E
Temperature
Keyed none
TC temperature
Measured
ET
Controlled/Meas none Keyed none
TC temperature
Measured
ET
HART
E
Controlled
ET RTD temperature
Measured
ET
Controlled/Meas none Keyed none
Simulated
ET
HART
E
RTD temperature
Measured
ET Frequency
Measured
E
Controlled
ET
Keyed none
Controlled/Meas none HART
E
Simulated
ET Value
Keyed none
Frequency
Measured
E Switch
Measured
E
Generated
ET
Keyed none
Value
Keyed none