EasyManua.ls Logo

BEAMEX MC5P - Frequency Generation

Default Icon
166 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
60
Startup and Basic Operation
Frequency Generation
The frequency generation terminals are located in the ET module.
Required settings Options/description
Quantity Freq.
Funct/Port ET: f(gen)
Also check the Unit and Waveform set-
tings.
MC5P generates sine or square wave with
given amplitude and frequency. Minimum
allowed frequency is 1 Hz, but the specifi-
cation is valid for frequencies above 40 Hz
only. When generating a square wave, the
output DC-level can be set to symmetric or
all positive.
The amplitude (V
pp
) setting range is: 0.01 V to 12 V.
Next…
Pulse Generation on page 61
Frequency Measurement on page 44
Special Generations on page 65
Calibration, see Part D.
ELECTRICAL MODULE
FOR MC5P
S/N
MULTIFUNCTION CALIBRATOR
MC5PE
?
ET
ELECTRICAL AND TEMPERATURE MODULE
FOR MC5P
S/N
Max 60 VDC/30 VAC
Max 60 VDC
Max 30 VAC
OUTPUT
SENSOR MEASURE & SIMULATE
INTERNAL REFERENCE JUNCTION
FOR THERMOCOUPLES
T/C, Low V
4-w meas
R, RTD
V, I,
3 & 4-w meas
Open
Close
Open
Close
T/C INT. RJ
T/C
WIRES
ONLY
(ENV), ENVIRONMENTAL
TEMPERATURE SENSOR
(EXT), EXTERNAL
PRESSURE MODULE
MEASURE
+24V
I
meas/sink
HART
®
2-w xmtr
V, ,
Low V
Com

Table of Contents