67
Warning!
Do not configure the range settings so that they exceed the
allowed input range of the connected instrument. MC5P deter-
mines the limits of the range settings based on the selected
quantity and port, not the connected instrument.
Ramping
The Ramping configuration windows shown
below can are opened as described in
chapter Opening the Step or Ramp Con-
figuration Window on page 65.
Configure the ramping settings according
to your needs. The following pictures are
examples of ramping configurations. The
graphs presented after the configuration
windows display the result of the settings.
zero) and a ramping with a predetermined
amount of repeats (max. value 65535).
To stop ramping press the function key
C/Stop Ramping. This applies for both a
continuous ramping (Repeats field set to
RAMPING
10.11.2003 16:10
Quantity
Port
Rise Time
Wait in 100%
Fall Time
Repeats
0=continuous
Range
Current
ET:I(gen)
1
1
2
2
8.0000
16.0000
s
mA
0 %
100 %
2
Wait in 0%
s
s
s
Stop
Ramping
Field StartCancel Edit
10.11.2003 16:11
Current
ET:I(gen)
5
0
0
0
8.0000
16.0000
s
mA
0 %
100 %
0 s
s
s
RAMPING
Quantity
Port
Rise Time
Wait in 100%
Fall Time
Repeats
0=continuous
Range
Wait in 0%
Stop
Ramping
Field StartCancel Edit
Current, mA
Time
, s
8
10
12
14
16
10-1 5 10 14
Start
Current, mA
Time, s
10-1 5 10 14
Start
8
10
12
14
16
Special Generations