EasyManua.ls Logo

BEAMEX MC5P - Using External Devices in Calibration Mode

Default Icon
166 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
122
Calibration
Using External Devices in Calibration Mode
In order to utilize external devices when calibrating an instrument,
the following things need to be set/available before the calibration
is started:
The required firmware option has to be installed in MC5P (and
when applicable, in the connected external device).
The Instrument’s Input Method field has to be set either to
Controlled or Controlled/Meas. About the settings:
Controlled: The input signal (pressure or temperature)
is measured using one of MC5P’s measurement input
ports.
Controlled/Meas: The input signal is measured by the
external device and communicated to MC5P.
In both cases: MC5P sends the set point values to the external
device via the communication cable connected to the AUX
connector
The Instrument’s Calibration Method field has to be set to
Automatic.
In the Instrument Window: Se-
lect the D/Menu and 7/Con-
troller Settings to select the
Controller type and activate
the communication.
Notes.
The Input Method option Con-
trolled is available for the follow-
ing Input Quantities: Pressure,
RTD temperature and TC tem-
perature.
The Input Method option Con-
trolled/Meas is available for the
following Input Quantities: Pres-
sure, Temperature, RTD Temperature and TC Temperature.
More information on how to create instruments into MC5P’s data-
base and how to edit the settings of an existing instrument is avail-
able in chapter Maintaining MC5P’s Instrument Database on page
124 and its subchapters.
Chapter Generating Signals Using External Devices on page 85
contains a description on how to utilize external devices in Basic
Mode and also information on controller settings.
11.11.2003 13:37
INSTRUMENT
112-TT-003.1
Edit Calibrate MENU
Intake Temperature on Feed 1
Cal. Points
5 ↑↓
Transfer Function
Linear
Method
Controlled
INPUT
0.00 ... 100.00 °C
Sensor Type
Pt100
α
385
Method
Measured
OUTPUT
4.0000 ... 20.0000 mA
Back
MENU
Close
Delete
Instrument
Controller
Settings
View
Instrument
Details
View
Calibration
Results

Table of Contents

Related product manuals