140
Appendix 1, User Guide for MC5 HART Option
For the Digital Output (PV), MC5P opens the Instrument Input Win-
dow for checking/editing the settings. Use B/Next Page Function
Key to check/edit the Instrument Output Window data.
Selecting the A/Trim Function Key opens the Calibration Windows
with the HART instrument’s trim menu opened. To change to Ana-
log Output (AO), use the menu option 4/Select Output.
Warning!
Do not alter any settings unless
you have the knowledge and are
authorized to trim the HART in-
strument.
Note.
Some of the menu options may be dis-
abled when trimming certain HART in-
struments. Refer to the HART
instrument’s manual for device depen-
dent information concerning the trim-
ming procedure.
11.11.2003 14:43
INSTRUMENT INPUT
Input Method
Measured
Unit
mbar
Range
0.000000
Pressure
Port 1
INT1C
Pressure type
Gauge
0 %
1000.0000
100 %
Trim
page
Next
Edit
11.11.2003 14:44
INSTRUMENT OUTPUT
Output Method
Range
4.000000
Current
Port
E
Unit
mA
0 %
20.000000
100 %
Transfer Function
Linear
Measured
Trim
page
Next
Edit
11.11.2003 14:53
Input
Error
1.00
Output
kPa
4.3362
-0.04 % of span
0.0433
+
0
-
0% 50% 100%
Pressure [P1:INT20C]
Current [E: Meas]
mA
gauge
Trim Lower
0% Range
Trim Upper
100% Range
Check Any
Value
Device
Information
Device
Settings
Select
Output
[AO]
End
HART
Adjustment
Analog Output (AO)
Digital Output (PV)