ATOMLAB 960 MEDICAL SPECTROMETER WIPE TESTING 8-5
Figure 8.5. Wipe Background Results screen.
7. At the Background Results screen the Isotopes section on the lower left allows the user to
select full spectrum, wide window or any other isotope that has been made available for this
location. Making a new selection changes the background results and the screen ROI to
reflect the background data for the specific isotope selected.
NOTE: You can see the Background results for each isotope by clicking on the isotope name.
8. Select <Accept> to save the background results currently displayed, or <Cancel> to return to
the Wipe Test screen.
9. After saving the background results, the Counting Wipe (For the Selected Location) screen is
displayed. Place that specific sample into the well counter. Select <Accept> to proceed with
the wipe count or <Cancel> to return to the Wipe Test screen.
10. The system now counts the wipe that is in the well counter. When the count is complete,
the Wipe Results screen for the selected wipe location just counted will be displayed.