EasyManua.ls Logo

Chroma A620028 - Page 111

Default Icon
238 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Manual Operation
3-63
62000H
DC Power Supply
+v
v
+
DVM
+sv
sv
Current Shunt
Electronic Load
+
+
Breaker
Figure 3-110
1. Calibrate the offset voltage of the ULTRASTAB SATURN before using
it for the current calibration. Press “
Null
” and the DVM panel digits
will reset and show “Math” for the current calibration.
2. Key in at least 5 digits for each calibration point to ensure the Power
Supply accuracy after calibration.
3.3.7.2.3 Calibration Procedure (Example: Model 62150H-600S)
1. In the CONFIG Setup page, press then press to enter the
CALIBRATION option (see section 3.3.7 above).
2. In the CALIBRATION page, press “ ” or turn the “Rotary” ( ) knob to set
CHOICE=2.
3. Press “ to display the current calibration options shown in Figure 3-111.
[ C U R R E N T M E A S C A L I B R A T I O N
]
R E M O V E A L L L O A D I N G F R O M O U T P U T T E R M I N A L
P R E S S
[
E N T E R
] W H E N R E A D Y
Figure 3-111
4. Open the Breaker to ensure the DC Power Supply has no load and press to
confirm.
5. It is important to reconnect the DC Power Supply to a current shunt whose rating is
closest to but still covers 2A. For the Prodigit 7550, use the 2A shunt directly.
6. The calibration option will display as shown in Figure 3-112. Press “ . The low
current range will be calibrated first. The system outputs a fixed voltage and then sets the
loading current of the Electronic LOAD to 1.5A. The cursor stops at position [1] as shown
in Figure 3-112. Enter the current read by the Current Shunt (DVM) and press “ ” to
confirm and wait for it to end.
7
ENTER
2
ENTER
ENTER
ENTER
ENTER

Table of Contents

Related product manuals