169
figure 4 b) 2- range: 10s.
(7)SpO
2
specification
Name Specifications
Detection range and
accuracy
0~300bpm , accuracy ±1bpm
Display Resolution 1%
Detection accuracy Within range of 90%~100%,measurement error should be±1% (No
state of motion);
Within range of 70%~90%, measurement error should be±2% (No
state of motion).
Upper and lower limit
of alarm preset and
accuracy
NELLCOR: Upper limit of alarm: 1%~100%
Lower limit of alarm: 0%~99%
Error of alarm should be setting value±1%
Simulate: Upper limit of alarm: 3%~100%
Lower limit of alarm: 0%~99%
Error of alarm should be setting value±1%
PR range and accuracy
Comen
range:20 - 254bpm
accuracy:± 3 bpm
Masimo
range:25 - 240 bpm
accuracy:± 3 bpm (no movement)
± 5 bpm(no movement)
Nellcor
range:20 - 300bpm
accuracy:20 - 250 bpm: ± 3 bpm
(8)NIBP specification
Name Specifications
Measuring
mode
Automatic oscillation
Measurement
range and
accuracy
Measurement
range for big
Systolic blood pressure
40-270mmHg
Diastolic blood
pressure
10-210mmHg
Mean blood pressure
20-230mmHg