169
figure 4 b) 2- range: 10s.
(7)SpO
2
specification
Name Specifications
Detection range and
accuracy
0~300bpm , accuracy ±1bpm
Display Resolution 1%
Detection accuracy Within range of 90%100%measurement error should be±1% (No
state of motion);
Within range of 70%90%, measurement error should be±2% (No
state of motion).
Upper and lower limit
of alarm preset and
accuracy
NELLCOR: Upper limit of alarm: 1%100%
Lower limit of alarm: 0%99%
Error of alarm should be setting value±1%
Simulate: Upper limit of alarm: 3%100%
Lower limit of alarm: 0%99%
Error of alarm should be setting value±1%
PR range and accuracy
Comen
range20 - 254bpm
accuracy:± 3 bpm
  Masimo
range25 - 240 bpm
accuracy:± 3 bpm (no movement)
± 5 bpm(no movement)
  Nellcor
range20 - 300bpm
accuracy20 - 250 bpm ± 3 bpm
8NIBP specification
Name Specifications
Measuring
mode
Automatic oscillation
Measurement
range and
accuracy
Measurement
range for big
Systolic blood pressure
40-270mmHg
Diastolic blood
pressure
10-210mmHg
Mean blood pressure
20-230mmHg