170
Measurement
range for small
Systolic blood pressure
40-200mmHg
diastolic pressure
10-150mmHg
Mean blood pressure
20-165mmHg
Accuracy
±5mmHg, when non-invasive BP exceeds the above
range, the monitor can still display normally, but
without accuracy
Range and
accuracy of
static pressure
Scope: should be 0mmHg300mmHg
accuracy
should be±3mmHg
Range and
tolerance of
overvoltage
protection
Big mode 300mmHg
Small mode 150mmHg
tolerance ±3mmHg
Range and
error of alarm
preset
Big
Systolic blood
pressure
Upper limit 46-270mmHg
Lower limit 40-264mmHg
Mean blood
pressure
Upper limit 26-230mmHg
Lower limit 20-224mmHg
diastolic
pressure
Upper limit 16-210mmHg
Lower limit 10-204mmHg
Small
Systolic blood
pressure
Upper limit 46-200mmHg
Lower limit 40-194mmHg
Mean blood
pressure
Upper limit26-165mmHg
Lower limit 20-159mmHg
diastolic
pressure
Upper limit 16-150mmHg
Lower limit 10-144mmHg
Error ±0.1kPa or±1mmHg (Both maximum)
NIBP
measurement
recall
2000 NIBP measurementdata
9TEMP specification
Name Specifications
Measurement range and
accuracy
Detection range 0℃~50
Measurement
error
±0.1
Alarm setting and
accuracy
Alarm setting
range
Upper limit of alarm should be 0.1℃~50.0
Lower limit of alarm should be 0℃~49.9
Alarm error ±0.1
Display Resolution
0.1°C