171
10Respiration specification
Name Specifications
Way Thoracic impedance method
Respiration rate
detection and accuracy
Measuring range
Big 6bpm-120bpm
Small 6bpm-150bpm
Accuracy ±1bpm
Accuracy and error of
preset alarm of
respiration rate
Big
Upper limit not
narrower than
8bpm-120bpm
Lower limit not
narrower than
6bpm-118bpm
Small
Upper limit not
narrower than
8pm-150bpm
Lower limit not
narrower than
6bpm-148bpm
Error ±1bpm
Range and error of
suffocation alarm
Scope: 10s40sand no alarm
11CO
2
specification
Name Specifications
CO
2
’s range of
measurement
0mmHg150mmHg, And air pressure is supplied by the mainframe.
CO
2
’s resolution 1mmHg or 0.1kPa or 0.1%
Gain
×0 .25 、×0.5、×1、×2、×4
Unit
mmHg、KPa
CO
2
’s accuracy
±2mmHg between 0mmHg 40mmHg
;
±5% between
41mmHg 70mmHg;
±8% between 71mmHg 100mmHg;
±10% between 101mmHg150mmHg
Atmospheric
pressure/altidude
400850mmHg/0 5029.2 M
O2 compensation
5100 mmHg
Range of alarm set and
error of alarm set
0~ 150mmHg/020 KPa
±1mmHg/±0.1 KPa
Wave speed 2.5mm/s、25 mm/s
Equilibrium gas N2O,indoor air ,Helium