(12)IBP Specifications
Name Specifications
Number of IBP channel 2 channels
Names of channel
pressure
Artery pressure(ART), pulmonary artery pressure(PA), central vein
pressure(CVP), right atrial pressure(RAP), left atrial pressure(LAP),
intracranial pressure(ICP), plus pressure(P1,P2).
Range and accuracy of
measurement
ART 0~40kPa(0~300mmHg)
PA
-0.8~16kPa(-6~120 mmHg)
CVP
-1.3~5.3kPa(-10~40mmHg)
RAP
-1.3~5.3kPa(-10~40mmHg)
LAP
-1.3~5.3kPa(-10~40mmHg)
ICP
-1.3~5.3kPa (-10~40mmHg)
P1,P2 -1.3~40kPa(-10~300mmHg)
error of IPB’s alarm set ±0.1 or ±1mmHg of the set value
pressure sensor
sensitivity: 5μV/V/mmHg
range of impedance: 300~3000Ω
(13) AG Specifications
Name
Specifications
AG measurement feature of infrared radiation absorption
AG resolution
CO2:1mmHg
awRR:1rpm
AG gas
CO2 , O2, N2O and one of five anesthetics (Enflurane, isoflurane,
sevoflurane, halothane, desflurane)
range and accuracy of
AG alarm set
Alarm
specification
range
step
accura-cy
EtCo2
upper limit (lower limit +
2)~76mmHg
lower limit 0~ (upper limit -
2)mmHg
1mmHg ±1mmHg
FiCo2
upper limit (lower limit +
2)~76mmHg
lower limit 0~ (upper limit -
2)mmHg
1mmHg ±1mmHg
AwRR
upper limit (lower limit +
2)~100rpm
lower limit 0~ (upper limit -
2)rpm
1rpm ±1rpm
EtO2
upper limit (lower limit +2)
20~100%
lower limit 18~ (upper limit
1% ±1%