172
12IBP Specifications
Name Specifications
Number of IBP channel 2 channels
Names of channel
pressure
Artery pressure(ART), pulmonary artery pressure(PA), central vein
pressure(CVP), right atrial pressure(RAP), left atrial pressure(LAP),
intracranial pressure(ICP), plus pressure(P1,P2).
Range and accuracy of
measurement
ART 040kPa0300mmHg
PA
-0.816kPa(-6120 mmHg)
CVP
-1.35.3kPa(-1040mmHg)
RAP
-1.35.3kPa(-1040mmHg
LAP
-1.35.3kPa(-1040mmHg)
ICP
-1.35.3kPa (-1040mmHg)
P1P2 -1.340kPa(-10300mmHg)
error of IPB’s alarm set ±0.1 or ±1mmHg of the set value
pressure sensor
sensitivity: 5μV/V/mmHg
range of impedance: 3003000
13 AG Specifications
Name
Specifications
AG measurement feature of infrared radiation absorption
AG resolution
CO21mmHg
awRR1rpm
AG gas
CO2 , O2, N2O and one of five anesthetics (Enflurane, isoflurane,
sevoflurane, halothane, desflurane)
range and accuracy of
AG alarm set
Alarm
specification
range
step
accura-cy
EtCo2
upper limit (lower limit
2)~76mmHg
lower limit 0~ (upper limit
2)mmHg
1mmHg ±1mmHg
FiCo2
upper limit (lower limit
2)~76mmHg
lower limit 0~ (upper limit
2)mmHg
1mmHg ±1mmHg
AwRR
upper limit (lower limit
2)~100rpm
lower limit 0~ (upper limit
2)rpm
1rpm ±1rpm
EtO2
upper limit (lower limit 2)
20~100%
lower limit 18~ (upper limit
1% ±1%