179
"Signal too weak "
Something wrong happens to the
measuring curve so that the system can
do measurement, analysis and
calculation.
After inspecting whether patient type is
set correctly, please check the connection
of all parts or replace a cuff. If the error
persists, please contact the manufacturer
for maintenance.
"Pressure out of scope "
Something wrong happens to the
measuring curve so that the system can
do measurement, analysis and
calculation.
Please check the connection of all parts
or replace a cuff. If the error persists,
please contact the manufacturer for
maintenance.
"Arms movement " The patient’s arm moves.
Please do measurement again after
checking the connection of all parts and
the patient's condition. If the error
persists, please contact the manufacturer
for maintenance.
"Overpressure protection "
Maybe the air path has once been
folded.
Please do measurement again after
checking whether the air path is smooth
and the patient’s condition. If the error
persists, please contact the manufacturer
for maintenance.
"Signal saturated "
Something wrong happens to the
measuring curve so that the system can
do measurement, analysis and
calculation.
Please do measurement again after
checking the connection of all parts and
the patient's condition. If the error
persists, please contact the manufacturer
for maintenance.
"Measurement time-out "
Something wrong happens to the
measuring curve so that the system can
do measurement, analysis and
calculation.
Please do measurement again after
checking the connection of all parts and
the patient's condition. If the error
persists, please contact the manufacturer
for maintenance.
"Cuff type wrong "
Maybe the cuff doesn’t accord with the
set patient type.
After inspecting whether patient type is
set correctly, Please check the connection
of all parts or replace a cuff. If the error
persists, please contact the manufacturer
for maintenance.
"Pump with air leakage " NIBP air path is leaked.
Please check the connection of all parts
or replace a cuff. If the error persists,
please contact the manufacturer for
maintenance.
"NIBP measurement failure "
Something wrong happens to the
measuring curve so that the system can
do measurement, analysis and
calculation.
Please do measurement again after
checking the connection of all parts and
the patient's condition. If the error
persists, please contact the manufacturer
for maintenance.
"NIBP system failure " Something wrong happens to the Please do measurement again after