"Recorder head too hot "
Maybe the continuous recording time is
too long.
After the recorder is fully cooled, do the
record output. If the fault still exists,
please contact the manufacturer for
maintenance.
"The position of the recorder
head wrong "
The paper-pressing handle of the
recorder doesn’t work.
Depress the paper-pressing handle of the
recorder.
"Recorder short of paper " There is no record paper in the recorder. Install the record paper.
"Recorder having paper jam " The recorder has a paper jam.
Place the recorder correct first and then
start recording.
"Recorder communication wrong
"
The recorder communication is not
normal.
In the record setting menu, start the
record-deletion function that can make
the upper machine and the lower machine
reconnected. If the fault still exists,
please contact the manufacturer for
maintenance.
"The serial port of recorder
wrong in communication”
"The paper position of recorder
wrong "
The paper is not placed correctly in the
recorder.
Re-place the corder paper rolls at the
correct location.
"Recorder unusable " The recorder can’t be communicated.
In the record setting menu, start the
record-deletion function that can make
the upper machine and the lower machine
reconnected. If the fault still exists,
please contact the manufacturer for
maintenance.
"NIBP initialization wrong "
NIBP initialization is wrong.
Select the resetting function in NIBP
menu. If the error persists, please contact
the manufacturer for maintenance.
"NIBP self-check wrong "
"NIBP resetting wrong "
Abnormal resetting happens when NIBP
is measured.
Please do measurement again after
checking the air path of NIBP and
whether there is air jam. If the error
persists, please contact the manufacturer
for maintenance.
"NIBP communication wrong "
Something wrong happens to the NIBP
communication.
Select the resetting function in NIBP
menu. If the error persists, please contact
the manufacturer for maintenance.
"Cuff too loose or not joined " NIBP cuff is not joined well. Please reconnect the NIBP cuff.
"Cuff gas tube leaked "
NIBP cuff is not joined well, or the air
path is leaked. .
Please check the connection of all parts
or replace a cuff. If the error persists,
please contact the manufacturer for
maintenance.
"Air pressure wrong"
Something wrong happens to the
measuring curve so that the system can
do measurement, analysis and
calculation.
Please check the connection of all parts
or replace a cuff. If the error persists,
please contact the manufacturer for
maintenance.