177
"Key wrong ";
" Keyboard wrong 1";
" Keyboard wrong 2";
"Network initialization wrong
(G.)"
Something wrong happens to the
network of the system so the system
can’t start network function.
Please contact the manufacturer for
maintenance.
" Network initialization wrong
(Ram)"
" Network initialization wrong
(Reg)"
" Network initialization wrong
(Mii)"
" Network initialization wrong
(Loop)"
" Network wrong (Run1)"
" Network wrong (Run2)"
" Network wrong (Run3)"
"5V voltage too high"
Something wrong happens to the power
supply of the system.
Please contact the manufacturer for
maintenance if such a prompt appears
frequently.
"5V voltage too low"
"Power system wrong 3"
" Power system wrong 4"
"12V voltage too high "
"12V voltage too low "
" Power system wrong 7"
" Power system wrong 8"
"3.3V12V voltage too high "
"3.3V voltage too low "
" Button battery voltage too
high"
Something wrong happens to the button
battery.
Please replace the battery. If the fault still
exists, please contact the manufacturer
for maintenance.
"Button battery voltage too low.
Please change it."
The button battery is short of charge or
it is not fixed or the connection of
button battery is loose.
"Recorder self-check wrong "
The system fails in connection with the
recorder module in self-check.
In the recorder-setting menu, start the
record-deletion function that can make
the upper machine and the lower machine
reconnected. If the fault still exists,
please contact the manufacturer for
maintenance.
"Recorder voltage too high"
The recorder module voltage fails.
Please contact the manufacturer for
maintenance.
" Recorder voltage too low"