176
"ECG interference too strong "
Stronger interference signal appear in
ECG signal.
Check the connection of ECG R-lead
cable, the current status of the patient and
whether more actions happen.
"XX module initialization wrong
X"
Error X appears when XX module is
initialized.
Restart the monitor for trial. If errors still
exist, please contact the manufacturer for
maintenance.
"XX module communication
stop"
XX module can’t communicate with the
main system normally.
"XX module communication
wrong "
XX module can’t communicate with the
main system normally.
XX means all parameter modules in the system such as ECG module, NIBP module, SPO2 module, IBP module and etc.
"XX alarm limit wrong "
The alarm limit of XX parameter is
changed accidently.
Contact the manufacturer for
maintenance.
"XX measurement out of scope "
The measured value of XX parameter
goes out of the range of measurement
by the system.
Contact the manufacturer for
maintenance.
XX means the names of all parameters in the system, such as HR, ST1, ST2, RR, SPO2, IBP, NIBP and etc.
"The real-time clock needs
resetting.”
If the system time is 2000-1-1, the
system will remind users that this time
is wrong.
Reset the system time. After resetting,
please restart the monitor for fear of
unnecessary time error being saved.
"The real-time clock doesn’t
exist "
No button batteries exist in the system
or the batteries have no charge.
Increase or change the button battery.
"System failure: software"
" System failure: cmos full"
" System failure: cmos err"
" System failure: fpga"
" System failure 2"
" System failure 3"
" System failure 4"
" System failure 5"
" System failure 6"
" System failure 7"
" System failure 8"
" System failure 9"
" System failure 10"
" System failure 11"
" System failure 12"
"Keyboard unusable "; The keys on the keyboard are unusable.
Please check whether the keys are held
on to artificially or other articles. If no
abnormal pressing is found, please
contact the manufacturer for
maintenance.
"Communication with keyboard
wrong ";
The keyboard fails, so it is unusable.
Please contact the manufacturer for
maintenance.