72
“High” means the most serious PVCs alarm.
(4)
ALM REC(Alarm Record): Select the “ON” and the recorder will output during PVCs
alarm.
(5)
ALM HI(Alarm Upper Limit): PVCs alarm is based upon the set alarm upper limit. The
alarm will happen when PVCs exceeds the upper limit.
The adjustable range for upper limit and lower limit of alarm as follows:
Name Max. Upper Limit Min. Lower Limit Single Adjustable Quantity
PVCs 10 1 1
PVCs Alarm Message and Prompt Message:
When the alarm record switch in the related menu is turned on, the physical alarm arising out of
that the parameters exceeds the alarm limit will make the recorder automatically output the alarm
parameter values and related waveforms.
The physical alarm and technical alarm that may happen in PVCs parameter measurement are
listed in the following table.
Physical alarms:
Prompt
Message
Causes Alarm Level
PVCs too high PVCs measured value is above the set alarm upper limit. Selectable by users
Technical alarms
prompt message Causes Alarm
Level
Solutions
PVCs alarm
limit is wrong.
Function
safety failure
High
Stop use of PVCs alarm and advise the biomedicine
engineer or our company’s servicemen.
(6)
ARR ALARM(ARR alarm setup): Set the Arrhythmia Alarm
①
ARR TYPE :
ASYSRTOLE
、
VFIB/VTAC
、
R ON T
、
VT>2
、
COUPLET
、
PVC
、
BIGEMINY
、
TRIGEMINY
、
TACHY
、
BRADY
、
PNC
、
PNP
、
MISSED BEATS.
②
ALM ON/OFF :ON ,OFF
③
ALM LEV :HIGH ,MED, LOW
④
ALM REC :ON/OFF
⑤
ARR ALM QUICK SETUP :
The alarm fully open, the alarm fully closed, the record
fully open, to record fully closed, the alarm level. The user can select the alarm fully
open various arrhythmia alarm is set to "On", select "Alarm full off various arrhythmia
alarm set to" Off ". Similarly, the record fully open "alarm record switch can be all set
to" On ", the The record fully closed alarm record switch can be all set to" Off ".