Table 9
Electrical connecons
Descripon
Mang connector descripon/external
supply rang
Internal supply
rang
Mains connecon Refer to Table: Electrical data for the
mains connector ed to each variant.
Connector is either:
Harng Han® K4/4-F nger safe 09 38 008
2703, 6-16 mm
2
ne stranded wire (VDE
0295 class 5, refer to Table: Wire assembly
according to VDE 0295), 8.9 mm maximum
insulaon diameter
or
Harng Han® Axial Screw module 100 A (2
o required), part number of mang half
suitable for 10‑25 mm
2
wire is 0914 002
2753 or 0914 002 2751 for 16‑35 mm
2
wire. Use ne stranded wire (VDE 0295
class 5, refer to Table: Wire assembly
according to VDE 0295).
or
Harng Han® Axial Screw module 200 A (3
o required), part number of mang half
suitable for 25‑40 mm
2
wire is 0914 001
2763 or 0914 001 2762 for 40‑70 mm
2
wire. Use ne stranded wire (VDE 0295
class 5, refer to Table: Wire assembly
according to VDE 0295).
Refer to Installaon on page 28 for the wiring diagram
PDT interface XLR type 5-way plug 24 V d.c. 0.2 A
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
0 V
24 V
Transmit data
Receive data
Not used
System interface XLR type 5-way plug 24 V d.c. 0.75 A*
Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
0 V
24 V
Transmit data
Receive data
Not used
Ethernet interface
Standard RJ45 type or Neutrik
®
EtherCon
®
RJ45
(IEEE802.3i
10 Base T Ethernet)
EMS interface XLR type 6-way plug
External emergency stop switch
Pin 1 - supply, Pin 2 - return**
24 V d.c. 100 mA
Page 23
M58800880_H - Technical data