Type A (ROTATION): 01
hex
Type B (ACCELERATION): 02
hex
Type C (BRAKE): 03
hex
Type D (FAIL N.O.): 04
hex
Type E (WARNING): 05
hex
Type F (not ROTATION): 06
hex
Type G (ACCEL./BRAKE): 07
hex
Table 29 - Parameter setting value
Example:
1. Input operation port: 01
hex
= I/O Remote (X4)
2. TMS Option: FF
hex
= DISABLE
4. Second Damage Limit Option: 00
hex
= ENABLE
5. First Damage Limit Warning: 00
hex
= ENABLE
6. Pump Runtime Over Warning: FF
hex
= DISABLE
7. Pump Runtime Over Warning hours: 000003E8
hex
= 1,000 (×100 hours)
8. Imbalance Warning: 00
hex
= ENABLE
9. Pump Overload Warning: FF
hex
= DISABLE
10. Pump Overload Warning Motor current: 03E8
hex
= 1,000 (×0.1%)
11. Pump Overload Warning Rotational speed: 0000
hex
= 0 (×0.1%)
12. Serial communication time out: 003C
hex
= 60 sec.
13. PROG OUT pin: 05
hex
= WARNING
14. PROG IN pin: 02
hex
= RESET/2
nd
SPEED EN