9 STORAGE AND DISPOSAL
9.1 Storage of the STP pump
When planning not to use the STP pump over a long period (more than a few months), follow the
precautions below:
1. Store the STP pump in a vertical position.
2. Close the inlet port of the STP pump and vacuum it using a backing pump.
3. Introduce dry N
2
gas or dry air from the outlet port or the purge port.
4. Close the outlet port and purge port.
5. If the STP pump has used water cooling, introduce compressed air from one side of the cooling
water port so that no water remains in the STP pump.
6. DO NOT store the STP pump in the following areas:
• Areas of high humidity (if it must be stored in a high humidity place, insulate it from outside and
use a dehumidifier.)
• Areas are prone to temperature extremes; high temperatures of more than 55 °C (131 °F) and
low temperatures of less than -25 °C (-13 °F).
• Areas where corrosive gases may exist
• Areas subjected to water/dampness
• Areas subjected to excessive dust
• Areas with insufficient ventilation
• Areas subjected to strong magnetic and electric fields or radiation
• Areas which will be subjected to direct sunlight
• Areas subjected to mist
• Areas subjected to electric noise and vibration
9.2 Disposal
WARNING
When disposing of the STP pump, exhaust gases inside the STP pump thoroughly.
Residual gases may result in an accident when disposing of the STP pump. If the
STP pump has been used with reactive or corrosive gasses, always clean thoroughly
before disposing of it to avoid any injury. Confirm the characteristics of gas to be
used, referring to the Safety Data Sheet (SDS) you obtain from the gas supplier.