EasyManua.ls Logo

Edwards STP-iXA4507 Series - 5.4.18 Readcondition

Edwards STP-iXA4507 Series
163 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
STP-iXA4507 series Turbomolecular Pump
MT-91E-001-0
Page 107
SERIAL COMMUNICATION PROTOCOL
5
Parameter
6
7
8
ASCII
F
F
0
0
0
0
0
3
E
8
0
0
HEX
46
46
30
30
30
30
30
33
45
38
30
30
Parameter
9
10
11
12
13
14
ASCII
F
F
0
3
E
8
0
0
0
0
0
0
3
C
0
5
0
2
HEX
46
46
30
33
45
38
30
30
30
30
30
30
33
43
30
35
30
32
Parameter
15
ASCII
*2
*2
*2
*2
*2
*2
*2
*2
*2
*2
*2
*2
*2
*2
*2
*2
*2
*2
HEX
*1 Assign the parameter value the reading data of ReadOptionFunc (?=) or FFFF32003CFF
hex
.
*2 Assign the parameter value the reading data of ReadOptionFunc (?=) or F
hex
.
5.4.18 ReadCondition
Function: Reading the pump model and damage point.
Transmission frame:
*1
PCSIM
Stx
0
0
1
?
{
Etx
LRC
SIMPC
Ack
PCSIM
*1
Ack
SIMPC
Stx
0
0
1
Sp
{
Parameter 1 to 4
Etx
LRC
*1 The HEX code of ASCII character ' { ' is "7B".
Parameter
Item
Data format
Remarks
1
Pump model
160-bits hexadecimal
coded ASCII
2
[System reservation]
32-bits hexadecimal
coded ASCII
3
Damage point
16-bits hexadecimal
coded ASCII
4
[System reservation]
64-bits hexadecimal
coded ASCII

Table of Contents