STP-iXA4507 series Turbomolecular Pump
MT-91E-001-0
Page 94
SERIAL COMMUNICATION PROTOCOL
PAG
E
94
5
Remote mode
Value
I/ O Remote (X4)
1
COM1 (EtherCAT)
2
COM2 (X2)
5
COM3 (X5 STP-LINK)
6
[System reservation]
3, 4
Table 27 - Remote mode
Example:
Remote mode setting: 01
hex
= I/O Remote
TMS function setting: 00
hex
= ENABLE
Parameter
1
2
3
ASCII
0
1
0
0
*1
*1
*1
*1
HEX
30
31
30
30
*1 System reservation
5.4.11 ReadEvents
Function: Reading the "Error Record"; it gets the most recent 10 errors that have been detected.
Transmission frame:
PC->SIM
Stx
0
0
1
?
g
Etx
LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC
Stx
0
0
1
Sp
g
Parameter 1 to 11
Etx
LRC
Parameter
Item
Data format
Remarks
1
The number of
"Error Record"
8-bits hexadecimal
coded ASCII
Up to 10 errors
2 to 11
Error Record 1 to
Error Record 10
8-bits hexadecimal
coded ASCII
*1
*1 The values corresponding to the error message are transmitted (refer to Table 24 and Table 25). The larger parameter
number means the more recent errors. When the number of errors being detected is under the maximum number, the
parameter value larger than the number of errors being detected is set to 0.