STP-iXA4507 series Turbomolecular Pump
MT-91E-001-0
Page 82
SERIAL COMMUNICATION PROTOCOL
PAG
E
82
5
5.4 Command specifications
5.4.1 Command list
Function
code
Command/Query Name
Function
?
D
ReadMeas
Reading the measured rotational speed
Sp
E
Command
Sending commands of START, STOP, RESET (These
commands are valid only when it is sent to the serial port
set as the input operation port.)
?
F
ReadFailMess
Reading the errors being detected
?
M
ReadModFonct
Reading the pump operation mode and the errors being
detected
?
V
ReadVersion
Reading the software version
?
c
ReadCounters
Reading serial number, running time, and number of
launch times
?
d
ReadSetPoint
Reading the setting values of the speed set point and the
TMS temperature
?
e
ReadMotorTemp
Reading the measured motor temperature
?
f
ReadStatus
Reading the various settings
(Remote mode, TMS function, and emergency vent valve).
?
g
ReadEvents
Reading the error records
Sp
h
SetSpeedSetPoint
Changing the speed set point
*1
?
h
ReadSpeedSetPoint
Reading the speed set point
?
m
ReadModFonctWithWarning
Reading the pump operation mode, the errors, and the
warnings being detected
?
[
ReadMeasValue
Reading the TMS temperature, motor temperature, motor
current, measured rotational speed, and control unit
temperature
?
=
ReadOptionFunc
Reading each setting value of items
Sp
=
SetOptionFunc
Changing each setting value of items
*1
?
{
ReadCondition
Reading pump model and damage point
?
}
ReadEventsWithTime
Reading the error record with detection time
Sp
0
SetOptions
Changing the optional function (Second speed option)
*1
?
0
ReadOptions
Reading optional function (Second speed option, TMS
temperature)
Table 22 - Command list
*1 There is an endurance limit of setting changes (no more than 24 times per day for about ten years). If this limit is
exceeded, it not only may become impossible to change the setting but can also cause pump failure. Do not make setting
changes in excess of this endurance limit.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards STP-iXA4507 Series and is the answer not in the manual?

Edwards STP-iXA4507 Series Specifications

General IconGeneral
BrandEdwards
ModelSTP-iXA4507 Series
CategoryWater Pump
LanguageEnglish

Related product manuals