14 Data Logger and Transmission of Data FLUXUS G60x
UMFLUXUS_G6V4-2EN, 2011-03-29 105
RS232
• default: 9600 bits/s, 8 data bits, even parity, 2 stop bits, protocol RTS/CTS (hardware,
handshake)
The transmission parameters of the RS232 interface can be changed:
14.2.5 Online Transmission of Data to a Terminal Program
• Start the terminal program.
• Enter the transmission parameters into the terminal program (see section 14.2.4). The
transmission parameters of the terminal program and of the transmitter have to be
identical.
• Select the program branch Output Options. Press ENTER.
• Select the channel for which the online transmission of data is to be activated. Press
ENTER until the menu item Serial Output is displayed.
• Set the storage rate (see section 14.1.2).
• Start the measurement. The measuring point number will be requested (see section
14.1.4).
14.2.6 Offline Transmission of Data to a Terminal Program
• Start the terminal program.
• Enter the transmission parameters into the terminal program (see section 14.2.4). The
transmission parameters of the terminal program and of the transmitter have to be
identical.
Enter HotCode 232-0- immediately after the transmitter has been switched on.
Set the transmission parameters in the 4 scroll lists. Press
EN-TER.
• baud: baud rate
• data: number of data bits
• par: parity
• st: number of stop bits
Select yes to activate the online transmission of data.
The measured values are transmitted during the
measurement.
Select Special Funct.\Print Meas.Val.. Press
ENTER.
baud<data par st
9600 8bit EVEN 2
Special Funct. ↕
Print Meas.Val.