GE HealthCare CONFIDENTIAL
Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
Chapter 3 LOGIQ Totus Setup 3 - 31
3-5-15 Logoff/Power shut down
You can use the SYSTEM - EXIT menu to log off, shut down, and exit.
1.) Press once on the ON/OFF button on the operator panel to display the SYSTEM - EXIT menu.
2.) Select Logoff to logoff the system, select Shutdown to shutdown the system.
Table 3-6 SYSTEM - EXIT Menu
Corresponding Graphic
The SYSTEM - EXIT menu, used when switching OFF the LOGIQ Totus, gives you these
choices:
Logoff:
Use this button to log off the current user.
The Ultrasound system remains ON and ready for a new user to log on.
If the Logoff button is dimmed, it indicates that no user is logged on to the LOGIQ Totus at the
moment.
Shutdown:
Use this button to shut down the Ultrasound system. The entire system will shut down. It is
recommended to perform a full shutdown at least once a week.
If the Shutdown button is dimmed, press the ON/OFF button or Alt+F10 to shut down the
LOGIQ Totus.
Cancel:
Use this button to exit from the SYSTEM - EXIT menu and return to the previous operation.
Exit (Only available when logged in as GE HealthCare Service with an SSA key):
Select this button when you want to exit to the Windows desktop.