Chapter 5 Components and Functions (Theory) 5 - 31
5-10-2 Power Supply Distribution
The main power module is located in the UBE-T assy, and main task is to isolate the scanner from the
on-site mains power system and to supply the various internal subsystems with DC power.
Power from the wall outlet (100 to 240 VAC, 50/60 Hz) is connected to the main power supply. The main
power supply (ACDC module) provides main P12V Live (Stand-by) 5V.
Converter (located on UCO-T,UAC-T,UBE-T) generates the needed voltages to the rest of the LOGIQ
Totus:
• Printer (12 Vdc)
• Main Monitor(12Vdc)
• Lower/Upper OPIO, Gelwarmer(12Vdc)
• HV (variable DC High voltage, +/-10V ~ +/-100V), THV(+/- 97V), on UAC-T
• P5V, M5V on UBE-T
• M5V on UAC-T
Figure 5-16 Power Distribution Diagram
The mains cord has plugs in both ends. A female plug connects to the scanner and a male plug to the
wall outlet.
User cannot detach the mains cord by regulation (fixed by Bracket).
Fuses are located inside the ACDC module.