Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
4 - 30 Section 4-3 - Functional Checks
3.) Then the system should be able to be unplugged and remain on for at least several seconds and
plugged back on without shutting down. A message may appear to user with a notification of being
in this mode.
4.) Verify that Scan is continued while the system is unplugged without shutdown.
5.) Reconnect AC power. Verify that the batteries begin charging via the battery status icon.
4-3-27 Shear Wave Elastography (Shear Elasto) functional checks
1.) Select a probe (C1-6-D).
2.) Select the ELASTO button.
3.) From the touch panel, select the Shear Wave button (if not already selected).
4.) Verify that the Shear Wave button is available.