GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-4 - Loading / Reloading / Upgrading the Software 8 - 11
8-4-7 Loading the base image and application software - full load
Follow this procedure to load the base image software and application software from the thumb drive.
NOTE: Lockout/tagout (LOTO) is not required as you will not manipulate any electronics.
NOTE: A Service key is not required to run this procedure.
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 and 8-2-7 "PPE Required During Service" on page 8-6.
Loading base image and application software
Table 8-5 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person
Approximately 1 hour and 30 minutes
to 2 hours and 30 minutes depending
on the amount of back-up/restore
activities (presets and patient archive)
needed prior to and after the
installation.
3 mm hex wrench
Software thumb drive
No PPE required
WARNINGWARNING
Before loading software, CONFIRM WITH THE CUSTOMER OR SITE ADMINISTRATOR
that the patient archive has been backed up. Failure to do so will cause permanent
DATA LOSS!
NOTICE
Before loading the base image load software, remove all media from USB ports and internal drives (for
example, DVD), and disconnect any portable hard drives. Failing to do so may interrupt the software
load process.