GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
8 - 112 Section 8-7 - Replacing OPIO
8-7-20 Replacement of the OPIO Mode Encoder Set-T
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-7-20-1 Preparations
1.) Remove the OPIO Top Cover Set-T as described in 8-7-2 "Replacement of the OPIO Top Cover
Set-T" on page 8-89
2.) Remove the UOPIO-T Assy as described in 8-7-10 "Replacement of the UOPIO-T Assy" on page
8-98
3.) Remove the T2 OPIO Upper Assy as described in 8-7-15 "Replacement of the T2 OPIO Upper
Assy" on page 8-105
4.) Remove the OPIO Laser Color Trackball with Cable as described in 8-7-11 "Replacement of the
Trackball assy-U_FLT20UFA-GJ" on page 8-101
5.) Remove OPIO Knob Set-T Assy as described in 8-7-16 "Replacement of the OPIO Knob Set-T" on
page 8-106
6.) Remove OPIO Lower Cover Set as described in 8-7-17 "Replacement of the OPIO Lower Cover
Set-T" on page 8-107
7.) Remove OPIO Button Set-T as described in 8-7-18 "Replacement of the OPIO Button Set-T and
the User Define Key Cap Set-T" on page 8-109
8.) Remove T2 OPIO Key Rubber as described in 8-7-19 "Replacement of the T2 OPIO Key Rubber"
on page 8-110
Table 8-56 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 15 minutes
Philips Screwdriver
Spanner 11
Cut resistant gloves
Knee Pad