GE HealthCare CONFIDENTIAL
DIRECTION 5936427, REV. 1 LOGIQ TOTUS Basic Service Manual
Section 8-7 - Replacing OPIO 8 - 93
8-7-6 Replacement of the OPIO Handle Cover Set-T
This table includes information specific to these instructions. For more information 8-2-6 "Tools needed
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
8-7-6-1 Preparations
1.) Remove the Probe Holder Set-T as described in 8-7-3 "Replacement of the Probe Holder Set-T" on
page 8-90
2.) Remove the OPIO Top Cover Set-T Assy as described in 8-7-2 "Replacement of the OPIO Top
Cover Set-T" on page 8-89
3.) Remove the UOPIO-T Assy as described in 8-7-10 "Replacement of the UOPIO-T Assy" on page
8-98
Table 8-42 Manpower, Time, Tools, and PPE
Manpower Time Tools PPE
One person Approximately 10 minutes
Flat Screwdriver
Philips Screwdriver
Cut resistant gloves
Knee Pad