Direction 5936427, Rev. 1 LOGIQ Totus Basic Service Manual
Chapter 4 Functional Checks 4 - 7
4-2-3 Power ON/Boot Up
Warnings
NOTE: Do not cycle the circuit breaker ON-OFF-ON in less than five seconds. When turning OFF the
circuit breaker, WAIT until the ON/OFF button is no longer lit. The LOGIQ Totus should de-
energize completely before turning the circuit breaker ON.
DANGER
ALWAYS CONNECT THE LOGIQ TOTUS TO A FIXED POWER SOCKET WHICH
HAS THE PROTECTIVE GROUNDING CONNECTOR.
DANGER
NEVER USE A THREE-TO-TWO PRONG ADAPTER; THIS DEFEATS THE
SAFETY GROUND.
DANGER
ENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND THAT THE
POWER PLUG IS THE PROPER HOSPITAL-GRADE TYPE (WHERE REQUIRED).
CAUTION
LOGIQ TOTUS REQUIRES ALL COVERS.
ONLY OPERATE THE LOGIQ TOTUS WHEN ALL OF THE BOARD COVERS AND FRAME
PANELS ARE SECURELY IN PLACE. THE COVERS ARE REQUIRED FOR SAFE OPERATION,
GOOD PERFORMANCE, AND COOLING PURPOSES.
CAUTION
Use only power supply cords, cables, and plugs provided by or designated by GE HealthCare.